51
<Val>
STR
32B length
Measured value at float format transmitted via
HEX characters (FF8100NN = error number
NN)
<ValStr>
STR
32B length
Measured value at string format (e.g. 12.8, n/a,
Error X)
<Unit>
STR
16B length
Unit of the channel at UTF-8
<Dec>
INT
0 - 10
Number of decimal places
<Alarm>
State of alarm at the channel (1 = alarm)
<_Al1>
INT
0, 1
Alarm 1 at channel
<_Al2>
INT
0, 1
Alarm 2 at channel
<AlarmLim>
Alarm limit at float format transmitted via HEX
characters.
IEEE 754 Float number 72.0442 is transmitted
as hexadecimal string 429016A0.
<_AlLim1>
STR
32B len
Alarm 1 limit for channel
<_AlLim2>
STR
32B len
Alarm 2 limit for channel
<AlarmMode>
Alarm mode:
0
alarm disabled
1
lower than limit
2
higher than limit
<_AlMode1>
INT
0
– 2
Alarm 1 mode
<_AlMode2>
INT
0
– 2
Alarm 2 mode
JSON field <AState>-<Reg> is a status of the device at moment when values
were sampled. Description of the status bits:
Bit
Description
bit0
– bit1
RSSI level indicator (0 = poor signal, 1 = sufficient, 2 = good, 3 = excellent)
bit2
– bit4
Unused
bit5
WLAN co-processor error (=1)
bit6
Internal hardware error (1 = Internal hardware error - RTC, EEPROM, LCD)
bit7
Device properly connected into Wifi (=1)
bit8
Optical LED active (=1)
bit9
Acoustic active (=1)
bit10
Unused
bit11
RTC time may not be valid - low voltage detected at RTC (=1)
bit12
Measurement error at one of channels (=1)
bit13
Configuration is not valid - configuration error (=1)
bit14
– bit15
Unused
Some bits at AState register can be deactivated at the configuration. Table