SECTION 7 APPENDIX I: MENUS, PROCESSED VARIABLES, AND GLOSSARY
______________________________________________________________________
7-3
Maintenance
Software Update
Default Settings
Language
View or Restore Factory Defaults
English
French
German
Portuguese
Spanish
Dutch
Italian
Swedish
Diagnostics
Impedance Checking
Diagnostic Codes
System Configuration
Advanced Diagnostics
On
/Off
On/
Off
Advanced
Diagnostics
DSC Self Test
Monitor Error Log
BISx Connection History
Clear Trend Data
7.3
Processed EEG Variables
Variable Name
Description
Range
Bispectral Index
(BIS)
The output from a multivariate discriminate
analysis that quantifies the overall bispectral
properties (frequency, power, and phase)
throughout the entire frequency range.
0 – 100
Signal Quality
Indicator (SQI)
A measure of the signal quality for the EEG
channel source(s) that is calculated based on
impedance data, artifact, and other variables.
Not affected by Suppression Ratio.
0 – 100
20 points per bar
EMG
The absolute power in the 70-110 Hz range. The
power value is reported in dB with respect to
0.0001
μ
V
2
. All the artifact detection is turned
off for this variable.
30 – 80 dB Trend
30 – 55 dB Bar
Graph
Suppression Ratio
(SR)
The percentage of epochs in the past 63 seconds
in which the EEG signal is considered
suppressed.
0 – 100 %