Start Procedure
2.5.1
Starting the Monitor for the First Time
To start the instrument for the first time, after it has been reset with the RESET button, or
after battery replacement:
1.
Attach one end of the power cord to the receptacle on the left side of the monitor.
2.
Plug the other end of the power cord into a properly grounded hospital-grade AC
power outlet. A yellow light illuminates to the right of the ON/Standby button.
3.
Press the ON/Standby button. The light changes to green and diagnostics tests run
to verify that the system is operating properly. (Do not connect or disconnect
equipment, or press keys until the monitor has completed its tests.) A beep indicates
that the tests are complete. If there is a problem, the system halts and an error
message appears. Error messages are explained in the Troubleshooting section of
this manual.
When not in use, the monitor should be placed in Standby mode. To
put the system in
Standby mode,
press and hold the ON/Standby button for two seconds before releasing.
The light will change from green to yellow. If the monitor is running on battery, the light will
go off completely.
Caution:
To completely remove power from the unit: put the monitor in Standby
mode, disconnect power cord from the power cord receptacle of the
monitor, then remove the battery from the monitor.
2.5.2
Starting the Monitor from Standby Mode
When the monitor is in Standby mode (yellow light), you may start it by pressing the
ON/Standby button. The light will change to green.
When not in use, the monitor should be placed in Standby mode. To
put the system in
Standby mode,
press and hold the ON/Standby button for two seconds before releasing.
The light will change from green to yellow. If the monitor is not connected to A/C power,
the light will go off completely.
Caution:
To completely remove power from the unit: put the monitor in Standby
mode, disconnect power cord from the power cord receptacle of the
monitor, then remove the battery from the monitor.