SECTION 5 PREVENTIVE MAINTENANCE, CARE AND CLEANING
______________________________________________________________________
5-4
5.2.4
Replacing the Battery
To replace the battery, you will need a Philips #2 screwdriver. Follow the instructions below:
1.
Unplug A/C line cord from the monitor.
2.
Lay monitor screen-side-down on a scratch-free work surface so that the
Battery/Power Supply cover is accessible.
3.
Remove 4 screws from the Battery/Power Supply cover and remove the cover. Note
the position of the battery cable.
4.
Squeeze the battery connector latch to disengage it from the back of the monitor and
remove the old battery.
5.
Lay the new battery in the recess with the wires at the top, and plug in the connector.
6.
Replace the cover and four screws (hand-tighten only) and reconnect the A/C power
cord.
5.2.5
Replacing the Power Supply
Figure 13 - Replacing the Power Supply
WARNING:
POWER SUPPLY IS INTERNALLY FUSED. REPLACE POWER
SUPPLY ONLY WITH ASPECT MEDICAL SYSTEMS BIS VIEW
POWER SUPPLY.
To replace the power supply, you will need a Philips #2 screwdriver. Follow the instructions
below: