Guidance and Manufacturer’s Declaration – electromagnetic immunity
The Triton™ Advanced Dental System® is intended for use in the electromagnetic environment specified below. The customer or the end user of the Triton™
Advanced Dental System® should assure it is used in such an environment.
immunity Test
en/iec 60601
Test level
compliance level
intended electromagnetic environment
Conducted RF
EN/IEC 61000-4-6
Radiated RF
EN/IEC 61000-4-3
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
Portable and mobile RF communications equipment should be used no
closer to any part of the Triton™ Advanced Dental System®, including
cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P 80MHz to 800 MHz
d = 2.3√P 800MHz to 2.5GHz
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
minimum separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey (a), should be less than the compliance level
in each frequency range (b).
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from objects, structures and
people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the Triton™ Advanced Dental System® is used
exceeds the applicable RF compliance level above, the Triton™ Advanced Dental System® should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the Triton™ Advanced Dental System®.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.