6.4  Troubleshooting the Self contained compressed Air and Suction Systems

Self-Contained – top four avoidable ServiCe iSSueS:

check these on your system to avoid future problems

#1 Over-filling the canister when flushing at end of day and pulling debris into the vacuum pump.   

Always follow instructions for cleaning the system and use limited quantities of water to flush the system appropriately.  
Do not flood it.

#2 Using bleach in the system to clean.    

Bleach will corrode internal parts of vacuum pump and requires rebuild kit (P/N 95-0068).  Do not use bleach to 
clean the system.  Only Use an Enzyme based suction line cleaner (P/N 95-0278).

#3 Odor build-up from not purging the suction canister daily or using enzyme based cleaner.  

(P/N 95-0278).  Clean daily and follow instructions.

#4 low Voltage Power Source.  

If operating at 120Volts/60Hz, the unit requires 115 - 125 volts for operation. Voltage levels below 115 will cause the 
motors to work harder, overheat, and may shut off temporarily until they cool down sufficiently.

Vacuum is low

-  Replace solids trap screens (P/N 95-0088 see section for complete list of part numbers available).

-  Crack in tubing or solids trap collector housing. Replace the housing if required. (P/N95-0247 or 95-0175 see section 

for complete list of part numbers available).

-  Piston cups in vacuum pump are worn. Use the Rebuild Kit to replace.  (P/N95-0068)

Vacuum Motor will not Start

-  The suction canister is full and the overflow switch is engaged.  Purge the suction canister completely.

-  The autoholder for the suction handpiece may be locked out. Remove the suction handpiece from the holder and 

ensure the lever is pointing down to the floor.

-  Start up capacitor may need replacement (P/N 95-0064 see section for full list of available part numbers).

-  Motor overload protection switch may have tripped. Let motor cool off for two hours and the compressor will 

automatically restart.

-  Low voltage power input. Verify 120 Volts or 230 Volts with multimeter. Correct lower voltage situation or switch to a 

dedicated outlet. For 120 Volts use optional power conditioner.

Suction canister will not Purge or empty Out

-  The pick-up tube in the canister is clogged. Remove and clean amalgam sediment bowl per instructions.

-  Peristaltic purge hose tubing is clogged or misaligned in the pump. Try massaging tubing to clear or adjust tubing in 

the pump. Replace tubing if necessary (P/N 95-0205).

Summary of Contents for Triton Advanced Dental System 2025M

Page 1: ...essory instruments from other manufacturers Please refer to the original manufacturer s instructions for operation and maintenance of those instruments Advanced Dental Systems Advanced Endodontic Systems Online www asimedical net Main 303 766 3646 Toll Free 800 566 9953 Triton Advanced Dental System Operation Service Manual Dental Operative Unit Models 2025M 2025M DV 2025M AR ...

Page 2: ... Evacuation System Optional 27 2 7 Self Contained Vacuum Pump 27 3 0 MAINTENANCE 28 3 1 Equipment Care and Inspection 28 3 2 Cleaning Disinfection Sterilization 28 3 3 Self Contained Models 30 4 0 STORAGE AND SHIPPING 32 4 1 Prepare for Storage 32 4 2 Shipping Preparation 32 5 0 LIMITED WARRANTY REV 0907 33 5 1 Specific Warranty Modifications 33 5 2 Specific Warranty Modifications for Self Contain...

Page 3: ...Electrical Wiring Compressor Vacuum Utilities Lower Compartment 62 8 7 Schematic Pressure Switch Air Compressor Lower Compartment 63 9 0 ILLUSTRATED PARTS LIST 64 9 1 Delivery System 64 9 2 Delivery Instrument Components 66 9 3 Air Water Syringe PN 20 0008 68 9 4 Closed Water System PN 80 0215 70 9 5 Delivery System Rear View 72 9 6 Cart System 74 9 7 Air Compressor Complete Assembly PN 85 0001 76...

Page 4: ...e periodically inspected and replaced if worn or damaged If an outside compressed air supply is used to power this unit the air supply must be regulated to 80 psi or below Excessive air pressure could cause certain components to rupture Warning Electrical Voltage This system is powered by high voltage electricity Like any other electrically powered device if it is not used properly it can cause el...

Page 5: ...e installed and put into service according to the EMC information stated below WARNING Use of portable phones or other portable or mobile radio frequency RF emitting equipment near the product may cause unexpected or adverse operation WARNING The use of accessories transducers and cables other than those supplied may result in increased emissions or decreased immunity performance of the product Th...

Page 6: ...use with 120volt 60Hz only 2025M DV Includes all features of 2025M above plus special HERM mounts on the air compressor and vacuum pump for added durability during transport and with dual voltage power input capability The dual voltage capability allows the system to be operated either from a 120volt 60Hz input or 120volt 50Hz input 2025M AR Includes all features of the 2025M DV above plus a power...

Page 7: ... 1 ASI PN 90 5147 Qty 2 Acteon Satelec Ultrasonic Handpiece Piezo ASI PN 90 5414 A Qty 4 NSK PANA Spray Lubricant Oil ASI PN 90 5714 Qty 1 KaVo Spray ASI PN 90 5715 Qty 1 box of 6 KaVo Quick Nozzle ASI PN 90 5715 N Qty 1 KaVo MULTIflex Nozzle ASI PN 90 5715 N2 Qty 1 Spray Nozzle PANA Spray ASI PN 90 5714N Qty 1 KaVo 7LP Lowspeed for Removable Attachments Ratio 2 7 1 ASI PN 90 5282 Qty 4 KaVo 181L ...

Page 8: ...ck and the flow of water by individual needle valves with knobs located on the front and side of the unit Electric instruments such as the ultrasonic handpiece and electric motor are controlled via the same control block However the compressed air that flows from the block is routed to an air electric switch that uses an air driven piston to close contacts when activated to thereby send an electri...

Page 9: ...n will turn off the motor This keeps the motor from running when the system is not in use The air compressor is mounted on special spring type isolators that allow it to float and thereby reduce vibration to the chassis A proprietary multi stage intake muffler design attenuates and muffles sound waves to dampen sound Air Tanks and Dry Clean Air The Triton system uses a special two stage air tank d...

Page 10: ...loat switch to prevent over filling of the canister Once the liquid rises to a pre set level the float switch will cut off the power to the vacuum pump Purge System The system uses an electric purge pump that can be turned on that will siphon the liquid from the holding tank and empty out though the purge hose at the back of the system At the end of the day s procedures the purge hose can be place...

Page 11: ...s it is rolled out of the case 5 Carefully grasp tubings and handpiece holder bag while removing cart Two people are required 6 Prepare to place handpieces in holders remove foot pedal from bag and place on the floor beside the cart 7 Unroll handpiece holder bag but be careful so as not to drop any handpieces 8 Lift handpiece holders so they are facing upwards Ensure levers on right side of holder...

Page 12: ...5M AR 12 65 2025M Rev B KaVo 25LCR KaVo Motor KaVo 18L NSK NL10Ti NSK NL9000 Electric Highspeed KaVo 7LP KaVo 68LH KaVo 10LP Air Lowspeed Optional Ultrasonic Air Water Syringe Saliva Ejector Curing Light High Volume Evacuator Fiber Optic Air Highspeed Fiber Optic Air Highspeed Lowspeed TO CART 1 4 1 Handpiece Setup ...

Page 13: ... equipped with a KaVo electric motor Two voltage selectors are located on top plate and the other two where the power cord enters the cart at the back of the cart Step 1 Ensure cart is unplugged from the power source Worktop is held in place with commercial grade Velcro Carefully remove top note that Velcro is very strong and will require a firm grasp Step 2 Locate the voltage switch located at th...

Page 14: ...ent may fail to operate when removed from the holder Verify Voltage Source The 2025M is dual voltage capable allowing it to operate either from 120 volt 60Hz North America typical or 230volt 50Hz Rest of World typical Severe damage can occur to the unit if it is set for the wrong voltage type Verify wall or generator voltage type Then verify correct voltage selection is complete for ALL THREE area...

Page 15: ...pump The motor will begin to run If it does not turn off and verify the waste tank has been purged The waste tank has a float switch that will turn off the vacuum pump if the container is filled Verify Water System Check and or fill water bottles appropriately and ensure the air pressure switch is turned on and that air pressure in the gauge reads approximately 30 35 psi Verify Handpiece Operation...

Page 16: ...ar All Systems Non retracting water system Electrical System 2025M 2025M DV 2025M AR Selectable by switches and Jumpers 115V 60Hz or 230V 50Hz Mains Power 9 15Amp at 115V or 4 55Amp at 230V Type B Compressor System 2025M 2025M DV 2025M AR Type Oil less piston Duty Cycle Intermittent duty Working Pressure 80 100psi 5 5 6 9bar Flow Rate 1 5cfm Sound Level 46dB Measured at 3 1m Air Reservoir 4 Liters...

Page 17: ...llowing are guidance and manufacturer s declarations regarding EMC for the Triton Advanced Dental System The Triton Advanced Dental System needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the following pages Warning This equipment is intended for use by healthcare professionals only As with all electrical medical eq...

Page 18: ...netic immunity of the Triton Advanced Dental System Guidance and Manufacturer s Declaration Electromagnetic Emissions The Triton Advanced Dental System is intended for use in the electromagnetic environment specified below The customer or the end user of the Triton Advanced Dental System should assure that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guid...

Page 19: ...vironment Surge EN IEC 61000 4 5 1kV differential mode line line 2kV common mode line earth 1kV differential mode line line 2kV common mode line earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines EN IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles ...

Page 20: ... the recommended minimum separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not appl...

Page 21: ...anced Dental System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Triton Advanced Dental System can help prevent electromagnetic interference by maintaining a minimum distance between the portable and mobile RF communications equipment transmitters and the Triton Advanced Dental System as recommended below acc...

Page 22: ...tles with clean demineralized water to a level below the screw threaded portion Re attach the bottles and hand tighten only 2 1 3 Operating the Dual Closed Clean Water System The closed clean water system features a selector valve that allows the operator to choose solution from bottle 1 or bottle 2 Turn the air pressure toggle switch to the ON position and ensure that the selector switch is flipp...

Page 23: ...ant spray generating a fog mist instead of just a stream of water To adjust coolant spray it is important to adjust the volume of chip air It may also be turned off to help prevent air from blowing into a treatment such as a surgical procedure Not all air driven high speed handpieces are ported for chip air so it is important to verify this option is available for the brand of handpiece being used...

Page 24: ...ndpieces can be attached Ensure that the handpiece you will be using has a gasket at the end and that it is in good condition to prevent potential air water leaks Place the handpiece onto the coupler and ensure that the tubes from the handpiece align in the connector and that the gasket on the back of the handpiece is seated against the connector Then firmly tighten the metal nut onto the handpiec...

Page 25: ...strument can be removed without having it activated by the rheostat To engage the lockout lift the lever to a horizontal position Figure 2 4 2 A To return the holder to normal operation point the lever back down to the floor Figure 2 4 2 B 2 4 3 Coolant Operation Adjusting Coolant Spray To activate the coolant water for a handpiece flip the toggle switch either located on the front of the system o...

Page 26: ...piece Note the power indication represents the power to the tip ranging from zero to 100 with 10 being 100 Many fine tips will break if they receive too much power Always start at a lower level and gradually increase to the desired power to avoid damaging tips Use the ultrasonic power setting table to determine the power range for the respective tip being used Ultrasonic Power settings table with ...

Page 27: ... solids trap 2 7 Self Contained Vacuum Pump The vacuum pump is activated when the Suction Vacuum On Off switch located on the rear of the unit is in the ON position and a suction handpiece is removed from the holder To use open the lever on the evacuation handpiece To maximize vacuum strength of the evacuation handpiece close the lever valve of the other evacuation handpiece whenever it is not in ...

Page 28: ...3 and request part number 95 0281 To use the cartridge can be attached to the pick up tube that hangs down in the water bottle by screwing it into the supplied fitting To replace unscrew the old cartridge and screw in the new one Place the Date Changed label on the outside of the clean water system as a reminder of when the next change is required Caution Do not use the DentaPure DP365 Cartridge f...

Page 29: ...d by others they must be properly sterilized before each use according to their manufacturer s instructions In addition to the disinfection of these parts a disposable barrier sleeve must be used to cover the air water syringe saliva ejector and high volume evacuator HVE The disposable barrier sleeve must be discarded and replaced between each patient Follow all manufacturer s instructions for ste...

Page 30: ... back of the unit above the waste purge tubing Flip the switch ON to purge the contents of the waste collection tank Caution The waste pump will remove the liquid under high pulsating pressure Therefore it is necessary to hold the hose by hand or ensure that it is firmly in place before pumping out the contents Note For hands free purging of the system the tubing can also be connected to a recepta...

Page 31: ... Pressure An air filter regulator see Figure 3 4 A controls system air pressure and is preset to 75 80 psi The air filter regulator is located inside the unit at the base of the aluminum air tanks To access remove the panel screws and open the panel Read the air gauge provided If the system pressure is not 75 80 psi lift the adjustment knob and turn clockwise to increase pressure or counter clockw...

Page 32: ...e cart 4 2 Shipping Preparation Caution The cart and trunk are very heavy Two people are required to safely pack up this system To pack the 2025M into the reusable ECS Shipping container trunk follow these instructions 1 Carefully place all of the handpiece tubing connections down into the individual slots of the handpiece holder bag Once all the connections are complete roll the bag and tie to se...

Page 33: ...ications The following items and or occurrences are not covered under the warranty Accidental Damage Theft including but not limited to the following Damage from accident or misuse including cuts tears and crushes to tubing cracks in suction solids trap filters housings broken casters etc Damage from use including scratches dings dents and deterioration to paint and overlays Theft or vandalism Dam...

Page 34: ...instruments as follows ASI s supplemental warranty does not extend the original term of the manufacturer s warranty include items not covered under the manufacturer s warranty or warranty problems caused by or resulting from misuse or accidental breakage Any installed instruments requiring repair will need to be removed and sent back to the respective manufacturer for inspection and repair It will...

Page 35: ...ure the correct adjustment stem on the control block portion for that handpiece has been opened Ensure that the exhaust air tubes and oil mist recovery jar are allowing the air to exhaust freely Diaphragm s in control block needs to be replaced P N 95 0215 see section for complete list of control block part numbers Verify that there is no kinked exhaust tubing to the oil mist recovery jar or any o...

Page 36: ...tyle P N 95 0135 that does not over retract Check the water on off toggle located either on the foot control or on the delivery system to see if it is working Remove the cover on the foot pedal from the toggle to see if the signal line could be kinked Remove the gray air signal line from the relay and activate the foot pedal and ensure that an air stream is coming out Check the water output side o...

Page 37: ...icant when lubricating instrumentation O rings Petroleum products will cause permanent damage to the o rings Spray mist occurs when depressing air valve immediately following water valve depression can be caused by worn nozzle assembly O rings Replace O rings Foot Control Sticking or Leaking Air If the disc is sticking or leaking air may need rebuild kit P N 95 0094 or grease the valve O rings wit...

Page 38: ...wise direction to fully open This ensures the Normally Open Air Electric Switch is receiving pressure from control block Check continuity output from Newtron Module 5 0 v AC 100 intensity Transformer 24v AC Terminal Block 120v AC Figure 6 2a Satelec Ultrasonic Notes Testing line continuity using a multimeter will determine broken wires in the tubing Figure 6 2a 6 2b 02 24 2014 S Technician Guide U...

Page 39: ...ctric Switch is receiving pressure from control block No Water through Ultrasonic Handpiece Water to Other Handpieces Make sure switch on foot pedal is in correct water position indicated by blue dot Check the adjustment needle valve is fully open on front of cart Check Ultrasonic Tip is secure and is correct type with a water port Some tips are made to run dry If in doubt remove the tip and run t...

Page 40: ...m is low Replace solids trap screens P N 95 0088 see section for complete list of part numbers available Crack in tubing or solids trap collector housing Replace the housing if required P N95 0247 or 95 0175 see section for complete list of part numbers available Piston cups in vacuum pump are worn Use the Rebuild Kit to replace P N95 0068 Vacuum Motor will Not Start The suction canister is full a...

Page 41: ...e stalling when attempting to start Problem is usually check valve allowing air back in and back pressuring the heads Remove check valve clean out and clean or replace Check Valve P N 95 0161 Start up capacitor may need replacement P N 95 0063 see section for full list of available part numbers Motor overload protection switch may have tripped Let motor cool off for two hours and the compressor wi...

Page 42: ...telephone To place an order please contact our Customer Support Department Online http asimedical net store Tel 303 766 3646 7 3 Repairing an Air Water and Air Only Syringe If there is water or air leakage around the button Replace the O ring on the button assembly To remove the button press it down and hold while using a paper clip or small wire to push the retaining pin out of the syringe head t...

Page 43: ...o the foot control and that the pressure has been exhausted After relieving the air pressure to the foot control turn it over and remove the outside screws from the bottom This will allow the dome and the retaining ring to be removed Once the dome and the retaining ring have been removed remove the valve stem and spring from the valve body Lubricate the stem with O ring grease and replace Ensure t...

Page 44: ...gh and low speed handpieces To adjust the drive air pressure for each handpiece perform the following steps 1 Remove the handpiece from its holder A handpiece must be connected to the tubing in order to register pressure on the gauge 2 Depress the foot control and read the handpiece pressure gauge on the front of the unit 3 Rotate the handpiece air adjustment clockwise to decrease or counter clock...

Page 45: ...tment to avoid vibrating against the air compressor as this can cause significant noise WARNING Electrical Shock The electrical terminal block is exposed Ensure that your HANDS and TOOLS are out of the unit before turning it on 7 8 1 Installing Compressor Rebuild Kit Replacement of Low Pressure Side Disassembly Refer to the exploded view parts list 9 6 during the repair procedures Note The air com...

Page 46: ...ntation for reassembly of the valves limiter and retainer mark if necessary 3 Remove the O ring and cylinder be sure to keep the same number and location of all shims 4 Remove the retainer screws retainer plate and piston cup 5 Discard the old cup cylinder retainer screws cylinder O ring head O ring valves and retainers these will be replaced by new parts in kit 7 8 4 Re assembly of parts for the ...

Page 47: ...te instructions or replace with a new one Reverse the above steps NOTE Ensure that the vacuum line tubing is securely pushed onto the fitting it should slide on at least You may wish to do a function check on the system before replacing the back panel WARNING Electrical Shock The electrical terminal block is exposed Ensure that your HANDS and TOOLS are out of the unit before turning it on 7 10 Rep...

Page 48: ...ed Wire Cutters Scissors WARNING Electrical Shock 1 Turn off main power and air supply to unit 2 Run Air Water syringe to relieve compressed air in system 3 Open the cover to the delivery system and locate the water relay valve near the control block Fig 7 11 A 4 Make note of the location for the blue water lines and the grey air signal line Clip all the tubings close to the barbs 5 Attach the blu...

Page 49: ... the hold down bracket to release the control bracket 5 Orient the control block with the heads of the socket cap screws facing up 6 Using a 7 64 Allen Wrench loosen all 13 screws from control block Turn the block over and carefully collect all the screws into your hand and then set aside 7 Lay the block flat and carefully separate the control block in the center open slowly so as not to lose any ...

Page 50: ...beled HP 1 Loosen the set screws with a small flathead screwdriver and remove the wires Fig 7 14 2 A 3 From the outside of the cart locate the black bulk head for highspeed handpiece and remove Fig 7 14 2 B Fig 7 14 2 C 4 On the new fiber optic handpiece cord there are 2 black wires that extend out of the tubing Make sure those wires are not crossed or tangled Thread wires through the small holes ...

Page 51: ...g 7 14 3 E 7 On the new ultrasonic cord remove the black plastic nut It is just like the one you removed in step 6 Fig 7 14 3 F 8 Feed new wires and tubing up through the existing hole in the baseplate 9 Once tubing is through the hole re string the black plastic nut onto tubing and tighten using Needle nose pliers like Step 6 Fig 7 14 3 G 10 Re connect the black plastic fitting to connect the wir...

Page 52: ...ystem bracket 5 Using 9 16 socket or wrench unscrew the nut holding existing valve in place then remove valve 6 Locate Blue waterline tubing on sides of existing valve and clip noting the locations of each tubing 7 Re attach tubing to new valve in same order as it was removed The bottom barb should go to 2 bottle head and the very top barb to the 1 bottle head The center barb is the output tube th...

Page 53: ...et and align the holes on top of pressure head with the holes in bracket with the screws then tighten the screws 9 Press the label back onto the top of the bracket 10 Fill bottles and attach Turn system on to pressurize bottles Test for leaks and proper water flow by running the Air Water syringe or Micro Irrigator handpiece 7 16 Compressed Air System 7 16 1 Remove Replace Air Compressor Tools Nee...

Page 54: ...remove the bowl and filter from the Master Regulator Fig 1 2 Using short stubby Phillips screwdriver remove the three screws from underneath the LEFT air tank Fig 2 3 Once screws have been removed firmly grasp the Regulator and pull to the left until it slides out of the air tanks Fig 3 4 Apply provided silicon grease to each side of regulator Fig 4 5 Slide the brass fitting into the Left air tank...

Page 55: ... 2 3 Using flathead screwdriver loosen hose clamp screw where tubing attaches to Suction connector on separator then pull tubing off the connector Fig 3 4 Repeat Step 3 on the Vacuum connector on separator 5 Locate purge pump connector Fig 4 There will be a small plastic ring on the inside of the connector that needs to be pressed in using a flathead screwdriver to disconnect Fig 5 6 Loosen Velcro...

Page 56: ... 56 65 2025M Rev B 11 To install new separator follow above instructions in reverse NOTE When attaching white wire to relay ensure that it is securely seated or the float switch may not function 12 To verify absence of leaks vacuum up some clean rinse water 7 17 2 Cleaning or Replacing Felt Dust Filter See section 7 16 3 ...

Page 57: ...CE 2 TO HANDPIECE 3 CONTROL BLOCK CONTROL BLOCK SIDE VIEW DRIVE EXHAUST CHIP WATER DRIVE EXHAUST CHIP DRIVE EXHAUST CHIP WATER WATER WATER IN SUPPLY AIR UNUSED POSITIONS DRIVE AIR FROM FOOT CONTROL OPTIONAL TOGGLE SWITCH IN DELIVERY UNIT AIR FOOT CONTROL AIR SUPPLY REGULATOR DRIVE AIR CHIP AIR SUPPLY AIR SUPPLY AIR WATER HOLDER TUBING EXHAUST AIR SIGNAL HANDPIECE ABBREVIATION CODES HS High Speed a...

Page 58: ...Tracer line from auto holder tubing attaches here Tracer line from auto holder tubing attaches here Water Drive Air Chip Air 5 Position Control Block Front Hand piece air adjustment knob 5 Position Single Control Block Side Water Chip Air Drive Air Pressure Gauge Main Air Output Ports to Handpieces Input Ports to Control Block Side To Flow Control Valve To handpiece on air electric switch From wat...

Page 59: ...L1 T1 PRESSURE SWITCH COMPRESSOR MOTOR VACUUM MOTOR 2 3 4 5 6 FILTER DRYER M LEVEL SWITCH TB2 RELAY SUCTION VACUUM SWITCH AIR COMPRESSOR SWITCH PURGE PUMP SWITCH POWER SWITCH N G L PURGE PUMP KCVO POWER SUPPLY FAN M INPUT 115V OR 230V JUMPER TRANSFORMER PHOTON POWER PACK 9VAC FHS 1 FHS 2 9V 1 2 3 115V 1 2 A B A B 115V 1 2 A B A B 115V JUMPER SETTINGS 120V SCHEMATIC TB1 P N 80 0175 1 L 2 N 3 A 4 A ...

Page 60: ...1 N L L2 T2 L1 T1 PRESSURE SWITCH COMPRESSOR MOTOR VACUUM MOTOR 2 3 4 5 6 FILTER DRYER M LEVEL SWITCH TB2 RELAY SUCTION VACUUM SWITCH AIR COMPRESSOR SWITCH PURGE PUMP SWITCH POWER SWITCH N G L PURGE PUMP KCVO POWER SUPPLY FAN M INPUT 115V OR 230V TRANSFORMER PHOTON POWER PACK 9VAC FHS 1 FHS 2 9V 1 2 3 230V 230V 1 2 A B A B 230V 1 2 A B A B 230V SCHEMATIC TB1 P N 80 0175 1 L 2 N 3 A 4 A 5 A 6 B 7 C...

Page 61: ... POWER SWITCH TRANSFORMER FHS 1 PHOTON POWER PACK FHS 2 AIR COMPRESSOR SWITCH SUCTION VACUUM SWITCH BLK WHT TERMINAL BLOCK TB1 P N 80 0175 BLU BRN BLU Blue BLU BLK BLK ORN BRN Blue TB2 2 TB2 1 Schematic 6 3 Line Cord BRN ORN ORN KCVO EHLS POWER SUPPLY TO FAN PURGE PUMP SWITCH PURGE PUMP WHT BLK WHT BRN BRN BLU BRN REFER TO DIAGRAMS TBD TBD FOR LARGER VIEW AND JUMPER CONFIGURATION JUMPER LOCATION U...

Page 62: ...acuum Pump Black to Relay Brown Brown Black to Pressure Switch White to Pressure Switch Air Compressor Switch Vacuum Switch Relay Float Switch Brown from Main Power Outlet Black to Air Compressor Switch Brown from Main Power Outlet Main Terminal Block TB1 Schematic 6 2 White to Float Switch Brown to Main Terminal Block Blue to Main Terminal Block Schematic Electrical Wiring Sub Lower Utilities Eff...

Page 63: ...M AR 63 65 2025M Rev B 8 0 Schematics 8 7 Schematic Pressure Switch Air Compressor Lower Compartment Schematic Pressure Switch Wiring Effective PRELIMINARY 3 2010 Pressure Switch Compressor Motor T1 L1 T2 L2 White Neutral White Neutral Black Line Black Line Power Input 120 230 VAC 115V 2A 2 2B 1A 1 1B RED ORG BLK BLU ...

Page 64: ...n www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 64 65 2025M Rev B 9 0 Illustrated Parts List 9 1 Dental Delivery System 1 2 3 8 8 8 31 9 13 4 4 5 16 30 32 11 24 19 15 8 17 23 6 7 7 18 25 26 12 22 21 20 10 20 29 28 14 22 ...

Page 65: ...nob Stainless Steel 5 16 15 0002 Hex Nut M S 18 8 S S 11841 4 17 95 0135 Valve Water Relay Stainless Steel 1 18 04 0275P Bracket Holder Bar Painted Colorado Beige 1 19 12 0083P Holder Bar Painted Colorado Beige 1 20 05 0037P Cube Holder Bar Painted Colorado Beige 2 21 15 0122 Screw Socket Set Screw S S 10 32 x 3 16 4 22 15 0057 Screw Phillips Panhead 10 32 X 3 8 SS 6 23 95 0204 Valve Needle High P...

Page 66: ...ASI The leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 66 65 2025M Rev B 9 2 Delivery System Instrument Components 10 3 7 9 5 4 6 8 2 1 ...

Page 67: ...ription Qty 1 Piezo Ultrasonic Generator Board 1 2 Power Supply Kavo Motor 1 3 Power Controller Kavo Motor 1 4 Transformer Dual Output 24 VAC 9VAC 1 5 Main Terminal Block 1 6 Power Supply Curing Light 1 7 Controller Fiber Optic System 1 8 Oil Collection Jar Handpiece Exhaust 1 9 Control Block 5 Position 1 10 Water Relay Valve 1 See Individual Part Descriptions for Detailed Part Number Information ...

Page 68: ...ASI The leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 68 65 2025M Rev B 9 3 Air Water Syringe PN 20 0008 Standard Syringe ...

Page 69: ...d Dental Systems 2025M 2025M DV 2025M AR 69 65 2025M Rev B 9 3a Parts List Air Water Syringe PN 20 0008 Figure ASI PN Description Qty 95 0235 Syringe Air Water No tubing 1 1 95 0010 Syringe Tip 1 2 95 0012 Syringe Repair Kit 1 3 95 0089 Syringe Button Repair Kit 1 Complete Assembly ...

Page 70: ...ASI The leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 70 65 2025M Rev B 9 4 Closed Water System Self Contained PN 80 0215 ...

Page 71: ...olorado Beige 1 2 95 0259 Valve On Off Relieving 1 3 95 0143 Valve Selector Routing 1 4 15 0080 Nut Hex 12mm 472 Nickel Plated 2 6 15 0139 Screw Phillips Flat Head 10 32 x 1 2 4 7 95 0266 Pressure Gauge 0 60 CWS System 1 8 15 0083 Nut Hex 4 40 S S 2 9 95 0216 Regulator Mini Inline 35 PSI 1 16 Barbs 1 10 95 0142 Pressure Head Clean Water Unit 2 11 95 0284 Tubing Water Line 1 4 Printed ASI P N 2 12 ...

Page 72: ...tion www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 72 65 2025M Rev B Refer to Illustration 9 1 Delivery System Rear View Photo Effective 3 26 10 10 10 14 15 16 12 3 7 13 9 5 4 6 8 2 1 11 9 5 Delivery System Rear View ...

Page 73: ...y Painted 1 4 04 0273P Door Interior CWS Painted 1 5 11 0183 Latch Knurled Cam Stainless Steel 1 6 04 0268P Panel Back Door Painted 1 7 80 0061 Cable Power Arm Cart Unit 1 8 04 0007P Bracket Cord Wrap Painted Colorado Beige 4 9 80 0110 Fan Self Contained Back Panel 1 10 11 0182 Latch Slotted Cam Stainless Steel 2 11 95 0306 Purge Hose Assembly 1 12 24 0001 Union Bulkhead 3 8 1 13 21 0093 Coupler F...

Page 74: ...ASI The leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 74 65 2025M Rev B 9 6 Cart System 10 12 3 7 9 5 4 6 8 2 1 11 ...

Page 75: ... 1 80 0188 Waste Container Amalgam Separator 1 2 42 0037 Relay Vacuum Pump Float Switch 1 3 95 0209 Sediment Bowl Amalgam Separator 1 4 95 0168 Vacuum Relief Valve 1 5 95 0167D Vacuum Pump 1 6 95 0178 Air Tank Regulator 1 7 95 0108 Check Valve 1 8 95 0159 Air Compressor 1 9 95 0205 Purge Hose Tubing 1 10 95 0293 Purge Pump Complete Assembly 1 11 95 0160 Pressure Switch 1 12 40 0067 Cooling Fan 1 ...

Page 76: ...eader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 76 65 2025M Rev B 9 7 Air Compressor Complete Assembly PN 80 0001 4 3 3a 3c 3b 1 1a 1b 2 5 7 6a 6b ...

Page 77: ... 20 S S 8 1b 15 00071 Screw Phillips Flathead 1 4 20 x 3 4 12 2 11 0222 HERM Isolator 4 3 80 0001D Compressor Manifold Dual Voltage 1 3a 95 0161 Valve Check 1 4 NPT Female Bronze Spring Loaded 1 3b 95 0160 Switch Pressure Air Comp 125 psi 1 3c 95 0228 Solenoid Valve Dual Voltage 1 4 40 0065 Compressor 1 2 HP With Filter Silencer Dual Voltage 1 5 95 0124 Capacitor 1 6a 95 0163 Filter Silencer 1 6b ...

Page 78: ...ASI The leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 78 65 2025M Rev B 9 8 Air Compressor Main Component PN 40 0019 ...

Page 79: ...pair Kit Head O Ring 2 2 Air Compressor Repair Kit Valve Limiter 1 3 Air Compressor Repair Kit Leaf Valve 2 4 Air Compressor Repair Kit Valve Retainer 3 5 Air Compressor Repair Kit Valve Screw 2 6 Air Compressor Repair Kit Cylinder O Ring 1 7 Air Compressor Repair Kit Cylinder 1 8 Air Compressor Repair Kit Piston Cup 1 9 Air Compressor Repair Kit Tube O Ring 4 10 Air Compressor Repair Kit Retainer...

Page 80: ...leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 80 65 2025M Rev B 9 9 Vacuum Pump Complete Assembly PN 85 0002D 1 1a 1b 1c 2 3 3a 5 4 6 6c 6a 6b ...

Page 81: ...solator 1 1a 15 0008 Finished Hex Nut 1 4 20 S S 8 1b 15 0016 Hex Head Bolt 1 4 20 x 3 4 S S 4 1c 15 0071 Screw Phillips Flathead 1 4 20x3 4 S S 12 2 11 0222 HERM Isolator 4 3 80 0036 Manifold Vacuum Pump 1 3a 95 0168 Valve Vacuum Relief 1 4 40 0066 Vacuum Pump 1 3 HP dual voltage 1 5 95 0125 Capacitor dual voltage 1 6 95 0169 Muffler Exhaust Vacuum Pump 1 6a 95 0163 Filter Silencer 1 6b 95 0093 F...

Page 82: ...ASI The leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 82 65 2025M Rev B 9 10 Vacuum Pump Main Component PN 40 0020 ...

Page 83: ...ain Component 1 1 Vacuum Pump Repair Kit Head O Ring 2 3 Vacuum Pump Repair Kit Leaf Valve 2 4 Vacuum Pump Repair Kit Valve Retainer 2 5 Vacuum Pump Repair Kit Valve Screw 2 6 Vacuum Pump Repair Kit Cylinder O Ring 1 7 Vacuum Pump Repair Kit Cylinder 2 8 Vacuum Pump Repair Kit Piston Cup 2 9 Vacuum Pump Repair Kit Tube O Ring 4 10 Vacuum Pump Repair Kit Retainer Screw 4 11 95 0117 Fan 1 12 95 0118...

Page 84: ...ASI The leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 84 65 2025M Rev B 9 11 Air Tank Dual Bracketed Complete Assembly PN 80 0192 1 2 ...

Page 85: ...iton Advanced Dental Systems 2025M 2025M DV 2025M AR 85 65 2025M Rev B 9 11a Parts List Air Tank Dual Bracketed Complete Assembly PN 80 0192 Figure ASI PN Description Qty 80 0192 Air Tank Dual Bracketed Complete Assembly 1 2 95 0178 Filter Regulator 1 Complete Assembly ...

Page 86: ... AR 86 65 2025M Rev B 9 12 Sediment Separator and Holding Tank PN 80 0188 9 10a Parts List Sediment Separator and Holding Tank PN 80 0188 Figure ASI PN Description Qty 80 0188 Holding Tank Complete 1 1 80 0187 Holding Tank Body 1 2 42 0022 Switch Liquid Level Polypropylene 1 3 95 0209 Sediment Bowl 1 Complete Assembly ...

Page 87: ... 87 65 2025M Rev B 9 13 Purge Pump PN 80 0214 9 11a Parts List Purge Pump PN 80 0214 Figure ASI PN Description Qty 95 0293 Pump Assembly Peristaltic 1 1 95 0307 Pump Peristaltic 3 Roller 1 6 mm Tubing 1 2 95 0308 Motor Gear 110 Volt 360 RPM 1 3 95 0205 Purge Hose Assembly Peristaltic Inner Purge Hose 1 Complete Assembly ...

Page 88: ...ASI The leader in Advanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 88 65 2025M Rev B 9 14 Foot Control Wet Dry PN 80 0112 ...

Page 89: ...ed Dental Systems 2025M 2025M DV 2025M AR 89 65 2025M Rev B 9 14a Parts List Foot Control PN 80 0112 Figure ASI PN Description Qty 90 0254 Foot Control Wet Dry 1 1 95 0240 Main Valve Repair Kit for Foot Control Wet Dry 1 2 95 0242 Switch Toggle Assembly for Foot Control Wet Dry 1 ...

Page 90: ... Ultrasonic Analog 95 0198 Module Satelec Ultrasonic SP Newtron 95 0100 O Ring Satelec Handpiece 95 0081 Potentiometer 95 0123 50 Tubing Handpiece Ultrasonic Satelec Specify serial number for proper strain relief 95 0123L 60 Tubing Handpiece Ultrasonic Satelec Specify serial number for proper strain relief 95 5414A Satelec Ultrasonic Handpiece 95 0137 Switch Air Electric Normally Open 95 0201 95 0...

Page 91: ... DV 2025M AR 91 65 2025M Rev B 9 16 KaVo ELECTROtorque TLC 9 16a Parts List KaVo ELECTROtorque TLC ASI PN Description 95 0309 Electrotorque TLC Basic Unit 95 0310 Display Circuit Board 95 0311 Membrane Keypad 95 0312 Motor Hose 95 0313 IntraLux Motor KL702 95 0314 Water Relay 95 0315 Transformer ELECTROtorque ...

Page 92: ...vanced Systems Integration www asimedical net Triton Advanced Dental Systems 2025M 2025M DV 2025M AR 92 65 2025M Rev B 7 1 2 3 4 8 12 13 14 15 16 9 10 11 5 6 9 17 Service Parts Kit PN 95 0107 Included with each model 2025M ...

Page 93: ...ncer replacement felt 2 3 95 0205 Tubing Peristaltic Pump 1 4 95 0089 Syringe Button Cartridge 2 5 95 0012 Syringe Repair Kit 1 6 95 0070 Repair Kit Leaf Valve Vacuum Pump 4 7 95 0127 Repair Kit Leaf Valve Compressor 4 8 95 0084 Switch Toggle Assembly for Foot Control Wet Dry 1 9 95 0156 Washer Pressure Head 4 10 95 0094 Foot Control Kit 1 11 95 0096 Kit Fittings with Grease 2 12 95 0224 Screwdriv...

Page 94: ...7 04 0275P Bracket Holder Bar Painted Colorado Beige 9 1 05 0037P Cube Holder Bar Painted Colorado Beige 9 1 09 0027C Worktop Classic Cart 9 2 11 0045 Hinge Black Panel 9 2 11 0051 Hinge Swing tip left 9 1 11 0052 Hinge Swing tip right 9 1 11 0182 Latch Slotted Cam Stainless Steel 9 2 11 0183 Latch Knurled Cam Stainless Steel 9 2 12 0071P Holder Bar 12 Sided 9 1 12 0071P Holder Bar 12 sided 9 2 13...

Page 95: ... 9 2 21 0104 Valve Water Relay Stainless Steel 9 1 21 0105 Pump Peristaltic 3 Roller 1 6 mm Tubing 9 11 21 0106 Tank Body Air Input 9 9 21 0107 Tank Body Air Output 9 9 21 0109 Regulator Mini Inline 35 PSI 1 16 Barbs 9 3 21 0118 Pressure Head Clean Water Unit 9 3 22 0102 Tubing Water Line 1 4 Printed ASI P N 9 3 22 0130 Purge Hose Tubing Assy 9 11 23 0049 Elbow 45 degree 1 4 Brass 9 9 23 0050 Conn...

Page 96: ...Arm Cart Unit 9 2 80 0109 Purge Hose Assembly 9 2 80 0110 Fan Self Contained Back Panel 9 2 80 0112 Foot Control Wet Dry 9 12 80 0143 Micro Air Pressure Gauge Assembly 9 1 80 0175 Terminal Bus 9 1 80 0185 Attenuator Air Compressor 9 5 80 0187 Amalgam Separator Body 9 10 80 0188 Amalgam Separator Complete 9 10 80 0192 Air Tank Dual Bracketed Complete Assembly 9 9 80 0212 Optional Accessory Outlet G...

Page 97: ... Kit Leaf Valve 9 6 95 0069 Air Compressor Repair Kit Piston Cup 9 6 95 0069 Air Compressor Repair Kit Retainer Screw 9 6 95 0069 Air Compressor Repair Kit Tube O Ring 9 6 95 0069 Air Compressor Repair Kit Valve Limiter 9 6 95 0069 Air Compressor Repair Kit Valve Retainer 9 6 95 0069 Air Compressor Repair Kit Valve Screw 9 6 95 0084 Switch Toggle Assembly for Foot Control Wet Dry 9 12 95 0089 Syri...

Page 98: ...needed 10 X Verify all door and drawer hinges and latches are undamaged lubricate if necessary 10 X Verify operation of water relay valve for opening and shutting Replace if necessary 10 X Replace gasket pucks in control block 90 X Replace gaskets in handpiece tubing bulkheads 30 X Replace handpiece tubing if worn or brittle 120 X Inspect display overlay replace if needed 10 X Foot Pedal Clean and...

Page 99: ...Check integrity of tubing and examine small wires for potential breakage 10 X Miscellaneous Electrical Inspect input wiring cord and internal wiring harnesses and ensure no damage or fraying 15 X Optional Suction Instruments and Accessories Clean and lubricate suction valve lever o ring and disconnect o ring P N 95 0085 grease 10 X Inspect clean and lubricate solids trap lid o ring 10 X Inspect al...

Reviews: