Store User Calibration
Perform the required method as described in the manual
using a standard of known concentration instead of
the water sample.
If the test result is
displayed, press [MODE], [Shift] + [4] [5] keys.
Confirm with [ ] key.
The display shows:
Pressing the arrow key [ ] once increases the
displayed result.
Pressing the arrow key [ ] once decreases the
displayed result.
Press keys until the displayed result corresponds to
the value of the standard.
Confirm with [ ] key.
Cancel user calibration by pressing [ESC] key.
The display shows:
100 Chlorine T
0.02-6 mg/l Cl2
0.90 mg/l free Cl2
<User calibration>
100 Chlorine T
0.02-6 mg/l Cl2
0.90 mg/l free Cl2
up: ↑, down: ↓
save:
Jus Factor
Saved