7.4 Manual Calibration:
The HC1000 offers a quick manual calibration mode, allowing the user to
easily calibrate the meter.
7.4.1
Record the cell constant value, which is clearly marked on a label
attached to the electrode (e.g., K=1.08).
7.4.2
Ensure the Cal indicator is showing, and press and hold the CAL
key for 3 seconds to enter setup menu, the display shows “CELL” option.
7.4.3
Press ENTER key to confirm, the meter enters the cell constant
setting mode.
7.4.4
Press ▲ or ▼ key to select “USER” option.
7.4.5
Press ENTER key to confirm, the meter enters User-Defined mode.
Enter
MI
Enter
Summary of Contents for WSL50 Pro
Page 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Page 4: ...Section 2 Kit Contents c c c c b a e f d ...
Page 7: ...Section 3 ...
Page 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Page 10: ...6 0 Guarantee and Assistance 32 ...
Page 12: ......
Page 16: ......
Page 21: ......
Page 33: ...Section 4 ...
Page 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Page 50: ...Section 5 HydroTest Photometer HT1000 ...
Page 86: ......
Page 136: ......
Page 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 193: ...NOTES ...