Press [Shift] + [1] keys to erase the method blank
and set the value back to factory calibration.
Fluoride Method 170
Press [MODE], [Shift] + [4] [0] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [3] keys.
The display shows:
1. Fill a clean vial (24 mm Ø) with exactly 10 ml of
deionised water, close tightly with cap.
2. Place the vial in the sample chamber, making
sure that the triangular marks on the vial and
instrument are aligned.
3. Press [ZERO] key.
The display shows:
4. Remove the vial from the sample chamber.
5. Add exactly 2 ml SPADNS reagent solution to
the water sample. Caution: vial is filled up to the
top!
6. Close the vial rightly with the cap and swirl
gently several times to mix the contents.
7. Place the vial in the sample chamber, making
sure that the triangular marks on the vial and
instrument are aligned.
8. Pres [TEST] key.
< Calibration >
1: M191 Ca-hardness 2
2: M191 reset 0 cali.
3: M170 Fluoride L
< Calibration >
170 Fluoride L
ZERO: deionised water
press ZERO
< Calibration >
170 Fluoride L
T1: 0 mg/l F
press TEST
Summary of Contents for WSL50 Pro
Page 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Page 4: ...Section 2 Kit Contents c c c c b a e f d ...
Page 7: ...Section 3 ...
Page 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Page 10: ...6 0 Guarantee and Assistance 32 ...
Page 12: ......
Page 16: ......
Page 21: ......
Page 33: ...Section 4 ...
Page 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Page 50: ...Section 5 HydroTest Photometer HT1000 ...
Page 86: ......
Page 136: ......
Page 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 193: ...NOTES ...