3.3.6 User defined Program (Fig 1.4)
This mode will run an incubation cycle at the user defined temperature for 24
hours. Adjusting the user defined temperature is described in section 3.3.9. The
24 hour timer can be altered to any setting between 0 and 24 hours. Adjusting
the incubation time is described in section 3.3.8.
Pressing and holding the middle switch for 5 seconds will start the incubation
cycle and the screen will change to the following:
The asterisk preceding the temperature will switch on and off indicating the
status of the heater.
To stop or pause the incubation cycle press and hold the middle switch again for
5 seconds. The screen will display the following:
Pressing and holding the middle switch will restart the cycle or pressing the left
switch will return to the menu function screen.
20.9°C
20.9°C
U->##°C MEM
OFF
*20.9°C
20.9°C
18:00 RUN! 24:00
*44.5°C
20.9°C
18:00 MEM 24:00
Summary of Contents for WSL50 Pro
Page 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Page 4: ...Section 2 Kit Contents c c c c b a e f d ...
Page 7: ...Section 3 ...
Page 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Page 10: ...6 0 Guarantee and Assistance 32 ...
Page 12: ......
Page 16: ......
Page 21: ......
Page 33: ...Section 4 ...
Page 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Page 50: ...Section 5 HydroTest Photometer HT1000 ...
Page 86: ......
Page 136: ......
Page 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 193: ...NOTES ...