Sample preparation:
Add one HT9 Ammonia No.1 tablet straight from the foil to the water
sample.
Crush the tablet using a clean stirring rod, until no large pieces are
visible.
Add one HT10 Ammonia No.2 tablet straight from the foil to the same
water sample.
Crush the tablet using a clean stirring rod, until no large pieces are
visible.
Close the vial tightly with the cap and swirl several times until the tablet
is dissolved.
Analysis:
Place the vial in the sample chamber, making sure that the marks on the
instrument and vial are aligned.
Press the Test key
Wait for a reaction period of 10 minutes
After the reaction period, the measurement starts and the result is
displayed in mg/l Ammonia as N.
Depress the cursor keys to cycle between N, NH
4
and NH
3
LOD/Tolerance