3.4.2
General
Finding
Possible Causes
Elimination
Test result
deviated from
the expected
Chemical species not as
required
Press arrow keys to select the
required chemical species
No
differentiation
e.g. for the
chlorine test
there is no
selection
between
differentiated,
free or total
Profi-Mode is switched
on
Switch Profi-Mode off with
mode 50
The pre-
programmed
countdown is
not displayed
Countdown is not
activated and/or the
Profi-Mode is activated
Switch the countdown on with
mode 13 and/or switch the
Profi-Mode off with mode 50
It seems that
a method is
not available
Method is not
activated in the user
method list
Activate the required method in
the user method list with mode
60