switched to automatic temperature compensation mode.
4.7 Manual Temperature Compensation:
4.7.1. DO NOT connect the temperature probe to meter.
4.7.2. Press and hold the MODE key for 3 seconds to enter temperature
setting mode.
4.7.3. Press ▲ or ▼ key to set the temperature of the sample.
4.7.4. Press ENTER key to confirm, the meter returns to measurement mode.
Setting is completed.
In the temperature s
etting mode, pressing and releasing the ▲ or ▼ key will
increase or decrease the temperature value by 0.1. Pressing and holding the
▲ or ▼ key will increase or decrease the temperature value by 1.
5.1 pH Calibration
The HC1000 allows up to a 5 point calibration in the pH mode. We
recommend performing at least a 2 point calibration for more accurate
measurement. The meter will automatically recognise and calibrate to
following standard buffer values.
USA Standard Buffer
Options
pH 1.68, 4.01, 7.00, 10.01, 12.45
NIST Standard Buffer
Options
pH 1.68, 4.01, 6.86, 9.18, 12.45
DIN Standard Buffer
Options
pH 1.09, 4.65, 6.79, 9.23, 12.75
Summary of Contents for WSL50 Pro
Page 1: ...AquaSafe WSL50 Pro Water Safety Laboratory Instruction Manual ...
Page 4: ...Section 2 Kit Contents c c c c b a e f d ...
Page 7: ...Section 3 ...
Page 8: ...AquaSafe Micro biological Incubator Instruction Manual ...
Page 10: ...6 0 Guarantee and Assistance 32 ...
Page 12: ......
Page 16: ......
Page 21: ......
Page 33: ...Section 4 ...
Page 34: ...AquaSafe Membrane Filtration Instruction Manual ...
Page 50: ...Section 5 HydroTest Photometer HT1000 ...
Page 86: ......
Page 136: ......
Page 142: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 193: ...NOTES ...