Place the vial in the sample chamber, making sure
that the marks on the instrument and vial are
aligned.
Press the Test key
Wait for a reaction period of 10 minutes
After the reaction period, the measurement starts
and the result is displayed in mg/l Nitrite.
LOD/Tolerance
The Lower LOD is 0.01 mg/l (10ppb), upper LOD is 0.5
mg/l (500ppb).
Tolerance: ± 0.005mg/l.
Notes
For best results, rinse vials thoroughly between with
deionised water. Ensure that the outside of the vials
are clean, dry and free from fingerprints. Always
handle vials by the lid where possible.
The tablets should be added direct from the foil
avoiding contact with hands or surfaces.