Switch the unit on using the power key.
Select the Ammonia method by depressing the cursor keys until
‘Ammonia T ’ is displayed or use the shortcut by pressing & holding shift
followed by 60
Release shift and press enter.
Blank analysis:
Ensure that the 10ml vial is clean.
Fill the vial with 10ml of the water sample.
Fit the cap and tighten.
Place the vial in the sample chamber, making sure that the arrow marks
on the instrument and vial are aligned.
Press the zero key
Wait until the following is displayed
Zero accepted
Prepare Test
Press TEST