13.6 Filling and Measurement Errors
13.6.1
Status Messages
Density condition
The "Density Condition" output field gives information about the current status of
measurement and may show the following messages:
Sample status icons
In the first column of the sample list, you find an icon for each sample that has
already been measured. The icon indicates the status of the sample:
13.6.2
Error Messages
Automatic bubble detection (FillingCheck
TM
)
If a bubble has been detected anywhere in the U-tube in real time, the "Density
Condition" output field shows "Filling Warning" or "Error: No Oscillation" for the
current sample and the diagnosis quick access button changes: Instead of the
green check mark, the yellow alarm sign is displayed.
Note: For complete transparency and traceability of your sample filling and
measurement process, check the filling visually by means of the built-in camera
(U-View
TM
, DMA 4500/5000 M) or the inspection window (DMA 4100 M) in
addition to the fully automatic FillingCheck
TM
.
temp. equilibration
Appears during temperature equilibration. In this stage
no predetermined result can be determined by the
instrument.
predetermined
Appears before the exact measuring temperature has
been reached and the instrument can determine a
predetermined result.
valid
Appears when the measuring temperature has been
reached and the measurement has been taken.
The sample was successfully measured.
Reduced precision due to e.g. a filling error.
No result due to a malfunction.