Appendix J: Firmware Versions
142
C76IB003EN-E
V2.00
10.09.2010
C76IB003EN-A
•
GUI performance improved
•
Data browser filter: only users and methods
used are available
•
More detailed information about firmware
module state during startup (new
initialization screen instead of online screen)
•
Safe mode operation after firmware errors
•
Sample list: ring buffer for 200 sample list
entries. Sample list entries older than 24
hours are automatically removed
•
Data memory can be set as ring buffer if
Audit Trail is deactivated
•
<Monitor> button to unfreeze the screen
moved to the quick access area
•
Allowed minimum step size in temperature
scans depends now on the equilibrium
setting of the density module
•
Keyboard and bar code settings: the sample
list can be edited with a bar code reader (tab
key as terminator)
•
<Redo> option after adjustments available
•
New air pump setting: Automatic stop when
the density value is stable
•
Audit Trail stores more detailed and
structured information on changes to the
system
•
New RS-232 commands: "set temperature"
and "help"
•
Printing a single sample on a list report is not
supported any more
•
Current API product group for API
calculation is now displayed in the calculator
V2.10
22.12.2010
C76IB003EN-B
New multiple measurements menu.
Firmware
version
Date of
release
Document
number
Comments