18 Communication with External PC and LIMS
C76IB003EN-E
111
RS-232 commands and responses
getdataunit
Gets unit data of the output quantities of the last
measured sample. For each output quantity the
unit is returned.
getdata
Returns the result values of the last
measurement only once.
finished
Returns the status of the measurement.
getrawdata
Returns the current raw data values for "Density
(not visc.-corr.)", "Density Temperature", "Set
Temperature" and the "Unique Sample ID". The
unique sample ID is "NaN" if no sample was
measured yet.
getmethodname
Returns the name and number of the method.
getid
Returns the serial number, instrument type,
firmware version and protocol version.
set temperature xx.xxx
Sets the set temperature of the current method
to the given value (xx.xxx in [°C]). This command
is only allowed, if no measurement is in
progress.
help
Returns a list of the available RS-232 interface
commands.
Command
Response
Description
start
or
start method_number
or
start method_number
magazine_position
measurement started
The command was accepted and
the measurement was started.
measurement already started
The measurement was already
started.
invalid method
The received start command did not
contain a method number or the
number could not be parsed.
simple mode not supported
An Xsample sample filling
equipment is active but the sample
list is in the simple mode which is
not supported. Change to the user-
defined sample list mode.
missing magazine position
An Xsample sample filling module is
active and the sample list is in the
user-defined mode but no magazine
position parameter was received.