17 Service Utilities
104
C76IB003EN-E
4. Select the content of your backup file in the column "Value" and tap <OK>.
5. Tap <Yes> to create your backup file.
The backup file is stored into the root directory of your USB flash drive.
17.2 Restoring Instrument Settings
For version 1.70 and higher, you can restore any backup file made with version
1.60 or higher.
1. Tap <Menu> and select
"Service > Restore Instrument Settings"
to open
the "Restore" dialog.
2. Tap <Details> to scan the content of the selected backup file.
3. Select the location of your backup (USB flash drive) and the "file name"
using the two drop-down boxes.
The backup file must be stored in the root directory of your USB flash drive.
4. Tap <Next>.
5. Select the settings to be restored to your instrument in the column "Value"
and tap <OK>.
6. Tap <Yes> to restore the settings.
7. Follow the instructions on the screen.
17.3 Updating the Firmware
Generally, it is only necessary to update the user interface firmware. This sets up
the instrument with the new features of the latest firmware version containing the
module firmware.
NOTICE
For instruments with firmware version 1.60 installed it is not possible to
restore backups made with older firmware versions.
NOTICE
When restoring the instrument settings, make sure that the backup file
corresponds with the instrument type.