17 Service Utilities
C76IB003EN-E
107
To change the log file configuration
In special cases your Anton Paar representative will ask you to change the
configuration of your log file in order to get more specific information about your
instrument.
1. Tap <Menu> and select
"Service > Update > Logging Configuration"
.
2. Insert a USB flash drive with the special configuration file delivered by your
Anton Paar representative into a USB interface.
3. Tap <OK>.
4. Save the instrument log file.
17.5 Viewing and Printing Live Raw Data
If the monitor mode is active, you get a live view of sensor signals including raw
data in this menu. You can export or print the live raw data.
After a measurement is finished, the live raw data of the main measuring unit (e.g.
"DMA Density") are frozen as well as the quantities on the main screen. To
unfreeze the values, tap the
button.
To view live raw data
•
Tap <Menu> and select
"Service > Live Raw Data"
.
DMA Density
Quantities not mentioned in the following table are described in Appendix F (e.g.
Density, Density Temperature, Density (not visc.-corr.) etc.).
dD Harmonic Value
Damping of the first harmonic.
Delta RHO Value
Difference between density calculated
from 1st harmonic and density calculated
from fundamental oscillation.
Density 1st Harmonic
Density calculated from the 1st harmonic
of the U-tube.
Density Progress
Indicator (0-100) for the progress of the
measurement.
e
Viscosity correction factor.
Fill Status
Status information about correct filling of
the U-tube. Values different from zero
indicate an incorrect filling.
RHO NC Harmonic Value
Not viscosity-corrected density calculated
form the 1st harmonic.
RHO17 Value
Viscosity-corrected density, valid if the
viscosity is less than 700 mPa·s.