59
Output of “Title Block” and “End Block”
When weighing values are recorded as the GMP, GLP report, “Title
Block” and “End Block” are added at the beginning and at the end
of a group of weight values.
Title Block
Step 1 In the weighing mode, press and hold the PRINT key
until
5tart
appears, and then release the key. The
scale outputs the “Title Block.” The scale automatically
returns to the weighing mode.
Step 2 Press the PRINT key or select the auto-print mode to
output the weight values.
End Block
Step 3
Press and hold the PRINT key until
recend
1
appears, and then release the key.
The scale outputs the “End Block.”
Step 4 The scale automatically returns to the weighing mode.
Press
and hold
To output the "
End block
"
Press
and hold
ZERO