16
Power on Tare Function:
When a tare weight (container) is placed on the weighing pan and the scale is turned on using the
ON/OFF key, the tare function is performed and a zero value is displayed as net value. Zero point
indicator and net indicator are displayed. The power on tare function can be used when the current
zero point is more than the threshold value
from the zero point of calibration.
Active function
Threshold value
Principal standard models
Approved models
Power on zero
function
±50 % of capacity
±10 % of capacity
Power on tare
function
+50 % of capacity
+10 % of capacity
7.3. Tare Function
The tare function is used when weighing an article in a container to display only the mass of the content
(net weight) by subtracting the mass of the container (tare weight).
CAUTION
The weighing range is reduced according to the value of the tare weight.
The current net value is reset to zero value when the ZERO key is pressed or the
display is turned off.
The storable preset tare value is within the minimum weighing range
7.3.1. How to Weigh a Tare Weight
Step 1 Place the container on the weighing pan.
Step 2 Wait for the stability indicator to be displayed, then press the TARE key.
The display becomes zero value.
Step 3 Place an article to be weighed into the container.
Wait for the stability indicator to be displayed, and then read the value.
Step 4 Remove all of the articles and the container from the weighing pan.
7.3.2. How to Input Value of a Tare Weight with Numerical Keys (Preset Tare)
Step 1 While pressing and holding the SET key, press the TARE key.
The scale displays a blank display or the stored tare value. A blank display means that the tare
value is zero (reset value). And
PT blinks.
Step 2 Input the preset tare value using the numerical keys.
Step 3 Press the ENTER key to store the new preset tare value.
The scale displays the net value as preset tare value subtracted from gross value.
Step 4 Place an article to be weighed into the container.