12. Calibration
Calibration is the function used to adjust the scale so that it can display the proper mass value.
Calibration must be performed in the following cases.
□
When the scale is initially installed
□
When the scale is moved to a location far away
□
When the ambient environment has greatly changed
□
At periodic calibration
12.1. Calibration Procedure
Step 1 Turn on the display. Open the rear cover of the display unit.
Locate the CAL switch inside.
Press the CAL switch to enter the calibration mode.
Then
Cal
is displayed.
Step 2 Select one of the following items to calibrate the scale.
ENTER : To calibrate using the calibration weight (Refer to “12.2. Calibration Using the Weight”).
MODE : To calibrate using the gravity acceleration value setting (Refer to “12.3. Setting the
Gravity Acceleration”).
12.2. Calibration Using the Weight
Calibration of the Zero Point
Step 1 Keep the display turned on for at least 30 minutes to warm up
the scale. Keep in mind the items of “4.1. Precautions for
Installing the Scale”.
Step 2 Press the ENTER key while
Cal
is displayed to display
Cal 0
.
Step 3
Confirm that nothing is placed on the weighing pan.
Wait for the stability indicator to be displayed.
Press the ENTER key.
The scale stores the current condition as the zero point.
Step 4 Calibration of the zero point is finished. To exit the
calibration mode at this step, press the CAL switch
.
The
display will then be turned off
CAL switch
STABLE
Stability indicator
Weighing mode
Note
The weighing pan shape
depends on the scale model.
Nothing on the
weighing pan
STABLE
Load cell Panel
CAL switch