31
12.3. Setting the Gravity Acceleration
When the scale is initially installed or is moved to another location,
the scale must be calibrated using the calibration weight. But if the
calibration weight cannot be prepared, change the gravity
acceleration of your scale to the value of the area where the scale
is used. See “Gravity Acceleration Map” of this manual.
Step 1 Press the MODE key while
Cal
is displayed.
Step 2 The scale enters the gravity acceleration correction mode.
Press the ENTER key to make the numerical value
blink and proceed to input mode.
Step 3 Set your local gravity acceleration using the numerical keypad.
Step 4 Press the ENTER key to store the new value.
The display returns to gravity acceleration display after
displaying
end
.
Step 5 Press the CAL switch to turn the display off.
Gravity acceleration setting is finished at this step.
Step 6 Press the ON/OFF key to turn the display on.
Confirm that weighing can be performed properly.
CAL switch
The power is automatically
turned off.
Enter using numerical
keypad
CAL switch