56
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration
executed
Calibration weight
Column for
signature
15.2. GMP, GLP Report
To print the GMP or GLP report to AD-8121B or AD-8127 printer, select the function setting
inf1 1
and set the AD-8121B to MODE 3 and the AD-8127 to dump print mode (DUMP).
To output the GMP, GLP report to a personal computer, select the function setting
inf1 2
.
Calibration Report
Step 1 Perform calibration according to “12.1. Procedure for the
Calibration”.
Step 2
end
appears when calibration is complete.
Step 3
gmp
appears and the calibration report is output.
Step 4
Cal
appears again. Remove the weight. Press the
ON/OFF key to turn the display OFF or press the CAL
switch.
AD-8121B, AD-8127 format
inf1 1
General format
inf1 2
: Space, ASCII 20h
CR: Carriage return, ASCII 0Dh
LF: Line feed, ASCII 0Ah
To output the results
A & D
MODEL HV-15KCWP
S/N 6A6123456
ID ABCDEF
DATE 2014/04/01
TIME 16:47:39
CALIBRATED(EXT.)
CAL.WEIGHT
+15.000 kg
SIGNATURE
`~~~~~~~~~~
A
~
&
~
D<CRLF>
MODEL
~
HV-15KCWP<CRLF>
S/N
~~~~
6A6123456<CRLF>
ID
~~~~~~~~
ABCDEF<CRLF>
DATE<CRLF>
<CRLF>
TIME<CRLF>
<CRLF>
CALIBRATED(EXT.)<CRLF>
CAL.WEIGHT<CRLF>
~~~~ ~
+15.000
~
kg<CRLF>
SIGNATURE<CRLF>
<CRLF>
<CRLF>
---------------
~
<CRLF>
<CRLF>
<CRLF>