43
Buzzer
Buzzer acoustic pressure level Approx.48 [dB/1m]
The buzzer is mounted on the electrical circuit board of the HVW-04CWP.
The buzzer can sound with LED (at upper side of the display) that is synchronized to a comparator
result. The use of the buzzer can specify at
of the function table.
Refer to “10.4. Buzzer Setting".
14.6. Communication Format
Data Format
S T
,
+
0
0
0
0
0
.
0
0
~
k
g C
R
L
F
Header Data Unit Terminator
There are 4 headers for the weighing data.
ST:
Stable weighing data
QT:
Stable counting data
US: Unstable weighing data
OL:
Out of weighing range
The data consists of 9 characters including the polarity and decimal point.
The weighing units are as follows.
means "Space (20h)".
~
k g: Weighing mode “kg”
~
l b:
Weighing mode “lb”
~
o z: Weighing mode “oz”
~
PC: Counting mode “pcs”
As a terminator, C
R
L
F
is always output. C
R
: 0Dh, L
F
: 0Ah
Data example
Weighing data “kg” (+)
S T
,
+ 0 0 1 2
.
3 4 5
~
k g C
R
L
F
Counting data “pcs” (+)
Q T
,
+ 0 0 0 1 2 3
4 5
~
P C C
R
L
F
Out of weighing range (+)
O L
,
+ 9 9 9 9 . 9 9 9
~
k g C
R
L
F
HV-CWP model
S T
,
+ 0 0 1 2
.
3 4
~
~
k g C
R
L
F
When the position of the readability changes depending
on the weighing range, the hidden digit is replaced to .
Example of tare function
When “tare data is output (
5ift 1
)” is specified in the parameter list, data is output.
Net
"kg" (+)
N
, + 0 0 1 2 . 3 4 5
k g C
R
L
F
Tare data
"kg"
T
, + 0 0 0 2 . 0 0 0
k g C
R
L
F
Example of preset tare function
Net
"kg" (+)
N
, + 0 0 1 2 . 3 4 5
k g C
R
L
F
Preset tare data
"kg"
P T , + 0 0 0 2 . 0 0 0
k g C
R
L
F