14
6.2. Keys
Keys
Description
ON/OFF Key
Shows or hides the display alternately.
Note: When the main power plug of the scale is connected to the receptacle,
the power is supplied to the electric circuit inside the scale.
ZERO Key
Based on the zero point weighed at the time of turning on the display (Power
on zero function), by pressing the ZERO key when the weight value is stable
within ±2% of the weighing capacity, the zero point indicator turns on as soon
as the display becomes zero.
At this time, if tare function is used, the tare value is deleted.
If accumulation data is displayed, the accumulation count and value are
deleted.
TARE Key
By pressing the TARE key when the weighing value is a positive and stable
value, the weight of the item on the weighing pan is subtracted as the weight
of a container (tare weight), and a zero value and both the zero point and net
indicators are displayed. (Tare function)
Note: The weighing range is reduced according to value of the tare weight.
M+ Key
Adds to the accumulated data.
SET Key
S and - for limit values in the comparator mode.
Press
and and
hold press
Use these keys to enter the preset tare setting mode.
Press
and and
hold press
Use these keys to display the accumulated results.
Press
and and
hold press
Use theses keys to set limit values for the comparator.
MODE Key
Switches the weighing unit to be displayed.
This key is used to select the items at each setting.
HOLD Key
Holds the display. (Refer to “13. Function” for details)
PRINT Key
Outputs the displayed weighing data. (Refer to “13. Function” for details)
Display off
Press
and and
hold press
Use these keys to enter the function mode.