Installation
|
3-9
Any screw terminals on the wiring card not described in this section
are reserved for future use.
This wiring section shows you how to make the following connections:
•
AC mains supply wiring to control unit
•
AC mains supply wiring to sensor
•
Control unit to sensor
•
Control unit to remote calibration unit
•
Control unit to alarm devices
•
Control unit to current output devices
•
Control unit to host computer (RS-485 communications)
In addition, this wiring section provides mandatory EMC grounding, shielding, and
noise protection requirements.
General Wiring and Conduit Requirements
This section describes general wiring and conduit requirements:
•
Sensor wiring conductors must be rated at a minimum of 80°C. All other wir-
ing conductor ratings should be for the minimum temperature required for the
equipment being connected to the analyzer, but not less than 60°C.
•
Use only the applicable NEMA approved conduit fittings or cable fittings to
maintain the NEMA rating for the Series 2000 controller or sensor enclosure.
If not using a conduit entry, leave the factory NEMA-approved plugs intact.
Never leave any holes unplugged.
•
Follow all applicable electrical codes for your location.
•
Follow proper grounding, shielding and noise protection practices as described
in this section.
•
Use twisted-pair cable with an overall braided shield, or twisted-pair cable in
rigid metal conduit for all sensor and signal wiring.
NOTE
Summary of Contents for CEM/Humox
Page 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Page 44: ...3 24 Thermox CEM Humox Analyzer ...
Page 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Page 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Page 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...