Calibration
|
5-11
The term zero value on the display is the value the system read during the
calibrate/verify operation, and the term zero gas is the setpoint value entered
using the Cal Gas Values option from the Calibrate key menu.
4. Press the Enter key. You are then prompted with the calibration data for the wet
oxygen cell.
The time when the calibration or verification was completed then appears on
the display:
End time hr:mn
End date mm/dd/yr
If the date or time has not been set, random characters will appear instead of
the time and date when the calibration or verification was completed. If this
occurs, reset the control unit’s time and date.