Controller/User Interface
|
4-23
where XX is the current filter value. Note that 100 equals no filtering, and 1
equals the highest possible filter value.
2. Type a number between 1 and 100, then press the Enter key. Note that you
should enter 01 to enter 1, 07 to enter 7, 70 for 70, etc....
The default value is 100 (no filtering).
Select Function
The Select Function menu option allows you to select what type of information the
analog output port should track. Choices are as follows:
Percent moisture
Percent oxygen (wet or dry)
Cell temperature (wet or dry)
Cell millivolts (wet or dry)
Thermocouple millivolts (T/C mv) (wet or dry)
1. Choose Select Function from the Analog Range menu after selecting an analog
output port.
2. Select the function you wish the selected analog output port to track (see bullets
above).
The Series 2000 control unit then prompts you with the Set Current Range menu
option where you choose the range of readings that the 0 to 20 mA or 4 to 20 mA
analog output represents (see the “Set Current Range” section in this chapter for
details).
Summary of Contents for CEM/Humox
Page 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Page 44: ...3 24 Thermox CEM Humox Analyzer ...
Page 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Page 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Page 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...