Maintenance and Troubleshooting
|
6-1
MAINTENANCE AND TROUBLESHOOTING
The operations in this chapter should only be performed by qualified
service personnel experienced with electrical safety techniques. There
are no operator serviceable components inside the CEM-Humox system.
Never service the controller or sensor unless power has been removed
from the controller and sensor, and the sensor has been allowed to
cool for at least one hour. Also, always use gloves when working on
the sensor.
This chapter describes Series 2000 control unit system and error messages. It also
provides troubleshooting assistance. System and error messages appear on the
fourth line of the display at three second intervals until the condition is corrected
or has ended. If there is more than one message, the messages will alternate on the
display at three second intervals.
These system and error messages are listed in this appendix alphabetically to make
them easier to locate.
System or error messages will not appear on the display when you are
using the control unit keys to access menu functions.
NOTE
Summary of Contents for CEM/Humox
Page 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Page 44: ...3 24 Thermox CEM Humox Analyzer ...
Page 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Page 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Page 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...