4-10 | Thermox CEM/Humox Analyzer
Wet Process Pressure
This menu option allows you to define your wet process pressure. It also allows
you to define whether your process is operating under a positive or negative pres
-
sure. Setting this value corrects wet oxygen measurements for the defined pressure.
1. Select Wet Process Press from the Setup key menu.
2. Select Positive Pressure or Negative Pressure.
You are then prompted to enter your system process pressure:
Pressure = xx.x PSIG
New Value?
3. Enter your process pressure (up to one decimal point is allowed - for example,
1.5 PSIG), then press the Enter key. If you select a positive pressure, a plus sign
(+) will appear before your value. If you select a negative pressure, a negative
sign (-) will appear before your value.
The process pressure you entered will now appear on the third display line.
4. If the correct pressure has been entered into the system, press the Enter key
without entering a new value to exit this menu option. If the pressure was incor-
rectly entered, you can reenter it at this point.
Summary of Contents for CEM/Humox
Page 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Page 44: ...3 24 Thermox CEM Humox Analyzer ...
Page 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Page 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Page 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...