4-20 | Thermox CEM/Humox Analyzer
where y is the analog output port, and xx is the present value that the current
output currently represents (for example, 10 % O2). Although we use output
port #1 for this example, the output port that will appear depends on the analog
port you selected.
2. Type the value to correspond to the 20 mA analog output.
You will then be prompted to define the value to correspond to the 0 or 4 mA
analog output:
Out # y 4 ma = xx
New Value?
where xx is the value that the current output currently represents.
3. Type the value to correspond to the 4 (or 0) mA analog output.