Installation
|
3-19
RS-485 Communications Connections
The Series 2000 control unit has the ability to communicate with other computers
that use the same protocol and have the same hardware interface. See the “Serial
Communications” Appendix in the back of this manual for information on the RS-
485 serial communications protocol.
2-Wire connections
The connections used for RS-485 2-wire communications on the wiring card are
labeled as follows:
4WTX+/2W+
4WTX-/2W-
Connect all the 2-wire pluses (+) and all the 2-wire minuses (-) in parallel with all
the devices in a daisy chain fashion, including all the control units on the network
and the host computer. In addition, on the host computer connect the receive and
transmit terminal block connections (see Figure 3-17). Be sure to use twisted-pair
cable for all connections.
2-Wire termination resistor
The Series 2000 control unit is equipped with a termination resistor that can be used
for the last control unit on the network. Switch 3 on SW1 of the display module
allows you to place a 120 ohm termination resistor into the RS-485 circuit:
Switch 3 of SW1 = ON (Termination Resistor In Circuit)
Summary of Contents for CEM/Humox
Page 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Page 44: ...3 24 Thermox CEM Humox Analyzer ...
Page 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Page 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Page 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Page 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...