Controller / User Interface | 4-61
Flow Switch Alarm
Allows you to set up the conditions for the
Low flow detected
alarm.
Alarm Type
Allows you to set the type of alarm that will
be displayed when a flow switching alarm is
detected. Options include:
0 = warning
(data is suspect; analyzer re-
quires service)
1 = fault
(data is invalid; serious failure
detected; sample system switches to
shut-in/off
mode)
Time Delay (sec)
The duration (seconds) that the Flow Switch
Alarm will be delayed before it is triggered.
Alarm Type
(Not available for Model 931 or Model 932 Analyzers.
)
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...