6-52 | Model 931S / Model 932S UV Analyzers
Decimal Alarm
Bit
Value
Type
Alarm Condition / Description and Suggested Corrective Action
6
64
Fault
f: Low flow detected
A low sample gas flow has been detected. This alarm will be delayed by the time
(seconds) entered in the
Flow Switch Alarm
Time Delay
field on the
Alarm
Setting and Stream Switch Parameters
dialog box.
Note: This alarm is used only by analyzers equipped with the optional flow
switch.
This alarm can be factory-set as a Warning or a Fault from the
Alarm
Type
field in the
Alarm Setting and Stream Switch Parameters
dialog
box, based on customer requirements. A Fault indicates a serious failure
has been detected and the data is not valid – the analyzer will switch
the sample system to
shut-in/off
mode. For a description of this alarm
if configured as a Warning, refer to the
w: Low flow detected
alarm
description earlier in this section.
Corrective Action:
•
Check the probe valves to ensure they are open.
•
Check the sample system to ensure there is adequate flow and there are no
plugs.
•
If your system uses an aspirator to move the sample gas, adjust the aspirator
air to ensure a proper flow rate, as indicated on analyzer drawings.
Also, check the aspirator air line for damage or bends.
•
Check the flow switch for plugs and/or for an open circuit.
Take appropriate safety precautions, open the Upper Enclosure, and:
•
Check the
Micro-Interface
board by simulating the low flow switch output.
Setup (tab)
Alarm/
Stream Switch
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...