6-6 | Model 931S / Model 932S UV Analyzers
Detecting a plug in the sample system:
This is best done by observing the analyzer’s measurement response
time immediately following a Zero. A typical response is in the range of
15–30 seconds. Observe your analyzer regularly to learn what a normal
response time is and keep records of response times after a Zero. Use
the recorded response times as a reference for detecting the formation of
plugs in the sample system (that is, longer response times).
The section of the sample system with the highest risk of plugging is
the sample inlet and the section of the sample inlet most prone to plug-
ging is the Sample Probe Valve. One way to confirm that a plug is in the
sample inlet is to cap the Vent Line (or, optionally, close the Vent isolation
valve) and Zero the analyzer. Plugs in the sample outlet (or vent side) of
the sample system are rare. If an aspirator is used to transport the sample
gas, a plug in the vent system can cause the aspirator drive air to flow
back into the Measuring Cell and, essentially, simulate a Zero Gas Flow or
Backpurge condition.
If the
Pressure Compensation
option is installed, observe the
Cell
Pres(sure)
reading on the
Cell/Miscellaneous
(compensation) parameters
dialog box, and if the reading approaches the Zero gas supply pressure, a
plug is present somewhere in the sample inlet.
To avoid damaging the pressure transducer, do not set the pressure
higher than the pressure transducer rating, either 210 KPAA (30
PSIA) or 0–700 KPAA (0–100 PSIA).
After confirming that the plug is somewhere in the sample inlet, take
all appropriate safety precautions and dismantle the sample system,
starting at the Sample Probe.
NOTE
!
CAUTION
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...