Controller / User Interface | 4-29
Default Cell Pres
Displays the default absolute pressure for the Measuring Cell
(mmHg). This value is used in place of the Measuring Cell pressure
when active pressure compensation is disabled. Enter a value to
place the pressure compensation into
fixed mode
(the approximate
Measuring Cell pressure).
Refer to the appropriate settings, below, for jumper selectable pres-
sure ranges for the corresponding 0–100 PSIA (0–700 kPaa), 0–30 PSIA
(0–210 kPaa), or 0–500 PSIA (0–3450 kPaa) Pressure Transducer.
If the 0–100 PSIA pressure transducer is being used, the jumpers
(
JP301
/
JP302
) on the
Customer I/O
board allow you to re-scale the
pressure range. See “0–100 PSIA Pressure Transducer” chart below.
0–100 PSIA Pressure Transducer Chart:
Jumper Settings, Customer I/O Board
PSIA
mmHg
JP301 JP302
0–100
0–5171.5
A
A
0–33
0–1706.6
A
B
33–66
1706.6–3412.2
B
B
66–100
3413.2–5171.5
C
B
While the 0–30 PSIA pressure transducer can not be re-scaled, its
jumper settings are listed in the “0–30 PSIA Pressure Transducer” chart
below.
0–30 PSIA Pressure Transducer Chart:
Jumper Settings, Customer I/O Board
PSIA
Pressure Range
JP301
JP302
0–30
0–1552 mmHg
A
A
If the 0–500 PSIA pressure transducer is being used, the jumpers
(
JP301
/
JP302
) on the
Customer I/O
board allow you to re-scale the
pressure range. The maximum allowable Cell Pressure is 200 psig
(1380 kPag). See “0–500 PSIA Pressure Transducer” chart below.
0–500 PSIA Pressure Transducer Chart (High Pressure/Low Range Kit):
Jumper Settings, Customer I/O Board
PSIA
Pressure Range
JP301
JP302
0–166.667
0–8619.16 mmHg
A
B
If changing the jumpers, be sure to make corresponding changes in
the S93XS System 200 Configurator Software.
NOTE
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...