4-46 | Model 931S / Model 932S UV Analyzers
Modbus Settings
The
Modbus
dialog box displays current communication parameters for
the analyzer you are communicating with. The parameters displayed here
are the results of settings in the analyzer.
Certain parameters need to be set up using the Configurator Software
to enable the Modbus port to establish communication with the Modbus
master.
Set up the following information:
Address
Enter the Modbus network address assigned to the analyzer.
Baud Rate
Select the baud rate of the Modbus network.
Stop Bits
Select the number of stop bits of the Modbus network.
Parity
Select the parity of the Modbus network.
Output Registers
button
Allows you to create a CSV (comma separated value) text file of all
of the Modbus Registers associated with the analyzer, and the value
(setting) for each register. The suggested filename will be predeter-
mined based on coded information from that analyzer (example,
m932-YY-932S-JJJJ-N). This file can then be imported and opened us-
ing MS Excel.
The field at the bottom of this dialog box (that is, RTU 9600 N 8 1 in Figure
4-15) represents the current settings.
Figure 4-15.
Modbus dialog box.
Setup (tab)
Modbus
See also “Digital
Communication” in Chapter 3.
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...