Controller / User Interface | 4-7
Configurator Software Troubleshooting
This section describes problems (alarms or “errors”) that may be encoun-
tered while attempting to establish communication between the analyzer
and the computer, or when starting the software. Communication error
messages are displayed in the
Status
field on the
General
tab and in the
Devices & Log
window. If this window is not displayed in the
AMETEK
Configurator
window (Figure 4-4), select
View
\
Devices & Log
or click
to display it.
If communication problems occur, review this section and perform the
corresponding corrective action to fix any problems.
Problem Encountered
“
Wrong Device
” is displayed on the
Devices & Log
window.
Corrective Action
Check the analyzer markings to ensure it is a Model 931S or Model 932S
Analyzer (the 93XS Configurator Software cannot communicate with other
analyzer models).
The
Communication
Protocol
/
Settings
information may be incorrect. Review
the settings in the
Communication Settings
dialog box (
Modbus Serial
or
Modbus TCP
) and verify or change them to the proper settings. The
Slave ID
(address) in this dialog box must also match the analyzer
Address
setting in
the
Modbus
dialog box.
Problem Encountered
“
Connected
” is displayed on the
Devices & Log
window, but the software is
not communicating with the analyzer.
Corrective Action
Enable
Live Data (General
tab
)
to establish communication.
Problem Encountered
“
DEMO Online
” is displayed in the
Status
field on the
General
tab. This mes-
sage will be displayed if
Live Data
is not selected (but
Device Enabled
is).
Corrective Action
Enable
Live Data
to establish communication.
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...