3-40 | Model 931S / Model 932S UV Analyzers
Start-Up Diagnostic Checklist
This checklist describes problems that may be encountered while power-
ing up the analyzer and/or preparing it for normal operation. If the ana-
lyzer experiences problems during power-up, review this checklist and
perform the corresponding corrective action to fix any problems.
Problem Encountered
The LEDs for various analyzer operations are not operating properly, as de-
scribed in “Powering Up the Analyzer” in this chapter.
Corrective Action
Take appropriate safety precautions and check for proper I/O wiring and
power connections.
If the LEDs are still not operating properly, reset the analyzer by cycling the
AC power off a minimum of 15 seconds before cycling it on again.
Problem Encountered
One or both source lamps are not firing.
Corrective Action
Take appropriate safety precautions and check for proper wiring connections.
If the wiring is okay, check the lamp-socket connections. See “Source Lamp
Replacement” in Chapter 6 to make adjustments to the lamps. Perform an
Auto-Setup if necessary to ensure the lamps are operating at peak efficiency.
Problem Encountered
Temperature zone values do not increase.
Corrective Action
Take appropriate safety precautions and:
• Check the fuses for the temperature zones to ensure they have not
blown. These fuses are located on the
AC Distribution
board in the
Lower Enclosure (see Figure 3-20).
• Check the Over-Temp Switch on the Optical Bench Assembly (see
Figure 3-2). If the plunger has popped out, press it to reset it.
• Check for proper wiring terminations for the appropriate temperature
zone.
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...