Maintenance and Troubleshooting | 6-7
Changing Out Replaceable Parts
The following sections discuss parts that should be replaced, as per the
“Analyzer Preventive Maintenance Schedule.”
Preventing leaks in the sample system is critical to proper analyzer
operation. If sample gas migrates into the Reflector Block or Optical
Bench (uncommon) due to a leak in the Measuring Cell, the optics
will become damaged and most likely require replacement. Most
leaks are preventable with regular cleaning and replacement of the
Measuring Cell o-rings.
IMPORTANT
Performing preventive maintenance on the analyzer requires using
the System 200 Configurator Software. Familiarize yourself with the
software before working on the analyzer. See Chapter 4 for details on
how to navigate through the software.
Measuring Cell Maintenance
This procedure discusses a Standard Range (40 cm) unheated Measuring
Cell (Figure 6-2). Other types of Measuring Cells are discussed in Manual
Supplements shipped with the analyzer. Refer to those Supplements for
specific cleaning information.
Replace/clean parts in the Measuring Cell:
• Every year.
• Any time an unscheduled cleaning is performed.
• If the analyzer responds slowly to a Zero when the sample system is
free of restrictions.
• If evidence of contamination is present in the sample tubing.
!
CAUTION
NOTE
See Chapter 7 for
spare part ordering numbers.
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...