Controller / User Interface | 4-45
UV Result Compensation
Species
**
Select the UV species the analyzer is measuring to compensate the
concentration measured by the sensor. Options include all UV spe-
cies measured by the analyzer and the calculated Custom result.
Select Disabled if no UV species compensation is required.
Factor
**
The factor used in the UV species compensation calculation.
Linearization Coefficient
(
Sensor1
and
Sensor3
screens)
Coefficients used in the linearization calculation applied to the con-
centration measurement from the sensor.
Cancel Calibration
button
Aborts the sensor manual span calibration sequence initiated by
ManSpan
button.
Display
During normal operation, displays the species being measured (top
line) and its real-time status (bottom line).
When an Auto-Setup is in progress, a message indicating the current
stage of the Auto-Setup is displayed on the top line (the concentration
value continues to be displayed on the bottom line).
During normal conditions, the text is Green; when there is a Warning
or Fault alarm present, the text is Red. If an alarm is present, refer to
the “Diagnostics and Troubleshooting” section in Chapter 6 for details
about the alarm and corrective action to take to reset the alarm.
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...