4-26 | Model 931S / Model 932S UV Analyzers
K Msr
**
The molar absorptivity of the gas species at the wavelengths used, for
the
Measure
PMT signal.
Model 931S Analyzers only.
K Ref
**
The molar absorptivity of the gas species at the wavelengths used, for
the
Reference
PMT signal.
Model 931S Analyzers only.
Key Filter
**
The position of the key filter, which gives the weakest signal for a
given source lamp current.
PMT Level (mV)
**
The Photomultiplier Tube (PMT) gain control signal, which is based
on the signal from the PMT when the key filter is in the light path.
After a successful Auto-Setup, the value should be 7200–7400 mV
DC
.
If the
PMT Level
is not in this range, adjust the
Lamp Max
and start
another Auto-Setup.
PMT Balance (mV)
**
Displays a secondary Photomultiplier Tube (PMT) gain control signal,
which adjusts the PMT gain to equalize the PMT signals when the
key filter is in the light path. After a successful Auto-Setup, the value
should be 2000–9000 mV
DC
.
Lamp Max (mV)
**
The voltage to which the lamp pulse current control signal is set to, for
the key filter location, during the Auto-Setup. To ensure a
PMT Level
of 7200–7400 mV
DC
after an Auto-Setup, the
Lamp Max
typical range is
4500–7800 mV.
Signal Max (mV)
**
Displays a value that is used by Auto-Setup to adjust the PMT sig-
nal (Reference or Measure PMT) with the highest gain. The PMT
signal with the highest gain is set to SigMax when each filter is in
the light path while the Auto-Setup is in progress. SigMax should be
8000–9000 mV.
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...