Calibration | 5-9
c. Span the third sensor (if used):
Connect the Span gas cylinder (for the species being measured by
Sensor3
, for example) to the analyzer.
Span the analyzer, as described in Step 2-a, then bleed down the
Span gas and disconnect it from the analyzer.
Continue with Step 3.
3. Change the
Interference Gas
values (
#1
and
#2
) back to their original
values, and click
Apply
.
4. Turn on Zero gas, and adjust it to original pressure, or as per recom-
mended pressure listed on the Final “As-Built” drawings.
5. Change the
Flow Control
to ‘
0
’ (Analyzer [
auto
] Control) and click
Apply
. Click
OK
to return to the
Setup
tab, then click
Save CFG
to EEPROM
to permanently save the changes to the analyzer’s
EEPROM.
If the Flow Control mode has been in a mode other than
auto (‘0’) for
longer than the
AZero Interval (hr) duration, and you attempt to
return it to
auto (by entering ‘0’ and then clicking the Apply button
next to
Flow Control text field), the mode will not switch until after
the
Auto-Zero Timer (AZ-0) duration, under Auto Calibration
Timers, elapses.
If the Flow Control mode is left in any mode other than ‘
0’ (auto),
the message “
Flow Control not in Auto” will be displayed in the
Operational Message window on the Status tab.
NOTE
Summary of Contents for 931S
Page 24: ...1 8 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 34: ...2 10 Model 931S Model 932S UV Analyzers...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 931S Model 932S UV Analyzers...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 931S Model 932S UV Analyzers...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 931S Model 932S UV Analyzers...
Page 96: ...3 50 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 170: ...4 74 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 184: ...5 14 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 237: ...Maintenance and Troubleshooting 6 53 Figure 6 14 AC Distribution board P N 100 1771...
Page 238: ...6 54 Model 931S Model 932S UV Analyzers Figure 6 15 Host Controller board P N 100 1757 SW100...
Page 239: ...Maintenance and Troubleshooting 6 55 Figure 6 16 Microcontroller board P N 100 1781 SW400...
Page 250: ...6 66 Model 931S Model 932S UV Analyzers This page intentionally left blank...
Page 263: ...Appendix A Drawings A 7 RS 232 Cable Communications Connections 300 9480...
Page 264: ...A 8 Model 931S Model 932S UV Analyzers Optical Bench Board 100 1662...
Page 265: ...Appendix A Drawings A 9 PMT Buffer Board 100 0140...
Page 268: ...S 2 Model 931S Model 932S UV Analyzers This page intentionally left blank...