Chapter 2: Installation, Page 12
M o d e l 6 8 2 T - H P U S E R M A N U A L A T E X Z o n e 2
2.3.2 Electrical
The incoming AC utilized for all analyzers should be “clean” AC, that is, AC with no transient
voltages on the line. No voltage dropout 10% below nominal and no voltage spikes 10% above
nominal VAC. AC voltages should be consistently maintained to 90% nominal VAC on the low end
and 110% nominal VAC on the high end. The AC voltage input can be 230 VAC single phase, 50 Hz,
1600 watts maximum power.
MAINS supply cords shall be rated for the maximum current for the equipment and the cable used
shall meet the requirements of IEC 60227 or IEC 60245. MAINS supply cords certified or approved
by a recognized testing authority are regarded as meeting this requirement.
The instrument must be connected to a protective earth ground.
Summary of Contents for 682T-HP
Page 1: ...Model 682T HP Sulfur Analyzer ATEX Zone 2 pzc User Manual 9000 223 VE REV C...
Page 25: ...Chapter 3 Operation Page 18 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Page 56: ...Chapter 5 Hardware Details Page 49 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Page 66: ...Chapter 5 Hardware Details Page 59 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Page 67: ...Chapter 5 Hardware Details Page 60 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Page 70: ...Chapter 5 Hardware Details Page 63 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Page 72: ...Chapter 5 Hardware Details Page 65 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Page 75: ...Chapter 5 Hardware Details Page 68 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Page 76: ...Chapter 5 Hardware Details Page 69 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...