Temperature control
198
Goldeye G/CL Technical Manual V4.1.1
LowerLimit
The sensor cooler operates below its lower power limit.
Under cold ambient conditions, the cooler may not be
required to keep the sensor temperature at the lowest
setpoint.
The image correction quality may also deteriorate during
this stage.
Red (flashing)
Alert
If the operating temperature of the camera is out of range,
this may cause an emergency shutdown. An emergency
shutdown is indicated by the LED continuously glowing red.
In this case, the color of the LED does not even change if
the camera cools down again.
If the internal camera temperature exceeds the defined
threshold temperature, the sensor and the critical
electronics of the sensor board shut down to prevent
overheating.
In this case, the camera can be restarted only manually by
switching the power off and on again. This is possible only
after the camera cooled down sufficiently.
Red (continuous)
LED color
flashing pattern
Status
Description
Table 123: Temperature Status LED (sheet 2 of 2)