3-8
N9927-90003
Tests and Adjustments
FieldFox Analyzer
Operational Check
5. Press
Mode
. If SA is available then continue below, otherwise the operational check is complete.
•
For this step Port 2 should be unconnected or connected to a short or load.
•
Press
SA
.
•
The measurement trace should be below the following limits:
Up to 4 GHz: limit =
−
45 dBm
Up to 6.5 GHz: limit =
−
42 dBm
Up to 9 GHz: limit =
−
39 dBm
Up to 14 GHz: limit =
−
33 dBm
Up to 18 GHz: limit =
−
28 dBm
Up to 26.5 GHz: limit =
−
18 dBm
•
The general slope of the trace should trend upward. A flat trace indicates an uncalibrated
condition that requires adjustments.
•
If a segment of the trace is offset by more than 10 dBm from the general slope of the trace, this
indicates a hardware failure that requires service.
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...