N9927-90003
7-27
FieldFox Analyzer
Repair and Replacement Procedures
Removing and Replacing the Measurement Group Assemblies
Removing and Replacing the Measurement Group Assemblies
Tools Required
•
T-8 TORX driver (set to 6 in-lb)
•
T-10 TORX driver (set to 9 in-lb)
•
Small flat blade screwdriver
•
5/16-inch open-end torque wrench (set to 10 in-lb)
•
9/16-inch open-end torque wrench (set to 21 in-lb)
•
. Refer to
“Miscellaneous Parts” on page 6-13
.
WARNING
Before beginning any repair, remove the battery and disconnect external power.
Removing All Measurement Group Assemblies as One Unit
1. Separate the front and rear cases. Refer to
“Separating the Front and Rear Cases” on page 7-10
2. Refer to
. Disconnect the A5 System board to A3 FPIB cable, W14, at the A3 FPIB
connector using the small flat blade screwdriver. If replacing this cable, disconnect both ends.
3. Lift the measurement group assemblies out of the front case and position them to the side of the front
case as shown.
4. Disconnect the A2 LCD to A5 System board cable, W11, at the A5 System board connector. It is
necessary to first remove the adhesive tape that is over the connector. To replace this cable, refer to
“Replacing the A2 LCD, the LCD Shield, and LCD Cables” on page 7-23
5. The measurement group assemblies are now completely separated from the analyzer.
6. To remove the I/O panel, refer to
“Replacing the I/O Side Panel, Doors, and Gasket” on page 7-14
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...