7-32
N9927-90003
Repair and Replacement Procedures
FieldFox Analyzer
Removing and Replacing the Measurement Group Assemblies
Replacing the A6 SOM Board
1. Separate the front and rear cases. Refer to
“Separating the Front and Rear Cases” on page 7-10
.
2. Remove all measurement group assemblies as one unit. Refer to
“Removing All Measurement Group
Assemblies as One Unit” on page 7-27
3. To replace the A6 SOM board:
a. Refer to
. Remove the RTC board. Refer to
“Replacing the Real Time Clock (RTC)
Battery Located on the A5 System Board Shield” on page 7-31
b. Remove the System board shield by removing the 19 screws (item 2). Discard the screws, new
ones are provided in the
c. Refer to
. Remove the four screws (item 2) that secure the A6 SOM board (item 1) to
the A5 System board. Discard these screws; new ones are provided in the
.
d. Unplug the A6 SOM board from the A5 System board.
e. Plug the new A6 SOM board into the A5 System board.
f. Install four new screws (item 2) from the
and tighten them in the
torque sequence indicated by the circled numbers (
➊➋➌➍
) in the illustration.
g. Replace the System board shield by replacing the 19 screws. Torque to 12 in-lbs.
h. Place the RTC board onto the System board shield and secure with the two screws. Torque to
9 in-lbs.
4. To remove the I/O panel, refer to
“Replacing the I/O Side Panel, Doors, and Gasket” on page 7-14
.
5. Perform the steps under
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...